High performance liquid chromatograph LC600
Technical Parameters:
1. LC-P600 high pressure constant flow pump
Basic technical indicators:
Flow rate range: 0.001~9.999ml/min to adjust the flow rate in 0.001ml/min steps.
Flow accuracy: RSD≤0.15%(1ml/min)
Flow accuracy: Ss≤2% (1ml/min water, 5~10MPa room temperature)
Maximum working pressure: 42MPa (0.001~9.999ml/min)
Pressure linearity and accuracy: display pressure error ≤0.5MPa (0~42MPa)
Pressure pulsation: ≤0.2MPa (flow rate 1ml/min, pressure 5~10MPa)
The tightness of the pump: the pressure is 42MPa, the time is 10min, and the pressure drop is ≤0.5MPa.
Dimensions: 450mm×300mm×160mm (length×width×height)
2. LC-UV600 UV detector
Basic technical indicators:
Wavelength range: 190~600nm
Spectral bandwidth: 8nm
Wavelength indication error: ≤2nm
Wavelength repeatability: ≤0.5nm
Baseline noise (peak-peak): ≤5×10-5AU (mobile phase methanol, 254nm)
Baseline drift (peak-peak): ≤5×10-4AU (mobile phase methanol, after stabilization, 254nm)
Minimum detection volume: 2×10-8g/ml (methanol solution of naphthalene)?
Dimensions: 450mm×300mm×160mm (length×width×height)